Double tilt 4 Electrode Transfer Holder

Q:What is the maximum heating temperature?
A:The maximum temperature is 1000°C.


Q:What is the sample size that can be placed on the MEMS chip?
A:The sample area is 10 μm × 100 μm.


Q:What is the maximum voltage that can be applied with the MEMS bias chip?
A:As of July 9, 2025, the MEMS bias chip can apply up to 10 V due to the controller l
   imitation. If an external power supply is used, the MEMS chip itself is rated for a    maximum of 20 V and 0.5 W.


Q:How is the temperature calibration of the MEMS heating chip performed?
A:Temperature calibration is carried out in batches. A representative chip from eac
    h batch is calibrated, and the resulting calibration file is applied to the other chi 
    ps in the same batch. The calibration is standardized using multiple experiment
    al data points based on the known properties of reference materials.