HATA-Holder System

tomography tem holder

3-axis orientation alignment Tomography
High efficiency EDS Analysis
Easy to link with other tools by using cartridges

Accurate 3D visualization (By courtesy of : Prof. Hata, Prof. Mitsuhara )

HPT-Cu

tomography tem holder

Austenitic steel

tomography tem holder

AL-Mg-Si

tomography tem holder

In general, the crystal orientation is aligned by two tilt axis(X tilt + Y tilt or rotation),
but X axis tilting, the crystal orientation is getting off axis.

Therefore, it is necessary for perfect alignment to apply not only Y tilting but also
specimen rotation, it can be tilted at a high angle on perfect aligned crystal
orientation, and a beautiful 3D construction is possible by this tem holder system.

Multi Function Structure : Virtual pivot

tomography tem holder

High detection efficiency structure for EDS analysis

Free form X-ray shadowing

tomography tem holder

EDS detection efficiency is improved by the geometry of HATA-Holder

tomography tem holder

High Resolution EDS analytical performance test : Column mapping of SrTiO3

tomography tem holder

Double tilt Cartridge Structure

tomography tem holder

Comparison Tomography between TEM and Atom probe Tomography

tomography tem holder

TEM Holder Specification

tomography tem holder
tomography tem holder

High-Angle Triple-Axis tem holder

Applicable instrument:
TFS Tecnai, Titan, JEOL HR pole piece type

Maximum X (α) tilt range +/-80°
( Limited by the Gonio stage tilt angle. )

Maximum Y (β) tilt range +/- 7.50°
Supports high-resolution(see right image)
Double tilt + cartridge structure

Sample rotation (IDS / Flexa)

Maximum Rotation range +/-5.00°
Option +/-100°
( Controlled by Mel-Build IDS or Flexa )

Minimum step control of angle : 0.01°
Illumination for specimen

tomography tem holder
tomography tem holder